control the conductivity of the cooling water is mandatory and is done through measuring instruments or conductivímetros. These instruments, we can even automate the process of purging and conduct operations depending on the extent of the conductivity continuously.
Currently, COYD, markets a wide range of equipment for this application, ranging from computers to measure individual panels complete and ready to connect to the installation.
It is a technologically advanced tool that allows making careful adjustments to the conductivity in industrial applications.
The LCD display is 15 mm high digits for
that can be read in areas with direct sunlight or
from a distance of 5 meters.
The box is made of plastic, particularly robust,
with an index of protection IP65.
The instrument is easy to use. The configuration
base is achieved through 6 microswitches
determining the reference temperature 20
or 25 ° C, the type of action of the relays, action
The alarm NA or NC and the protection of
programming and calibration functions.
The measuring probe is calibrated in a very
Simple.
The measurement is compensated in temperature
manual or automatic with three-PT100
threads.
The temperature is viewable with the push ° C.
Technical Data:
DIMENSIONS: 144 * 144 * 103 mm.
PLAN FOR FIXING: 136 * 136 mm +2 mm max.
INTERNAL DIMENSIONS: 90 mm.
DEGREE OF PROTECTION: IP65 front IP20 rear.
SECTION OF CABLE: 4mm2, or 2.4 mm diameter.
FOOD: Universal 22 ... 265V 50-60 Hz.
CONSUMPTION: 3W.
SET POINTS: 2 Independent configurable relays.
MAXIMUM LOAD ON RELAY: 3A 250VAC.
TEMPERATURE OF WORK: 0 ... 50 ° C.
Storage temperature: -10 ... +70 ° C.
RANGES OF ACTION AND PRECISION: 2 ± 0.01 / 5 ± 0.01 / 20 ± 0.01 / 50 ± 0.1 / 200 ± 0.1 / 500 ± 0.3 / 1,000 ± 2 / 2.000 ± 2 / 5,000 ± 5 / 10,000 ± 200 / 20,000 ± 200.
CONSTANT FOR CELL: 0.1 / 0.2 / 1 cm -1.
CAL 0: ± 10% F.S.
CAL GAIN: ± 30%.
TEMPERATURE COMPENSATION: manual / automatic 0 ... 125 ° C.
REGULATION ON THE SET POINT: ON / OFF - TEMPORIZED.
Hysteresis ON THE SET: Programmable.
DELAY ON THE SET 2: 0 ... 99 minutes.
TRANSMITTED OUTPUT: 4 ... 20 mA, resolution 200 points, 400 ohm load.
Measure:
K = 10 - Range: 0 ... 500 mS
Scales: 2-5 - 20 - 50 - 200 - 500
K = 5 - Range: 0 ... 1000 mS
Scales: 20 - 50 - 200 - 500 - 1,000
* K = 1 - Range: 0. . . 20,000 uS
Scales: 2,000 - 5,000 - 10,000 - 20,000
* The scales of measurement 0 ... 10,000 ... 20,000 uS and 0 uS are achieved only using conductivity probe
Model KTC CTK-1-1-G
| Model |
Feature |
Measures |
| CD75A |
Instrument wall IP65 |
P 144 x 144 x = 90 |
| CD75B |
instrument panel DIN |
P 144 x 144 x = 90 |
| CD75C |
Instrument panel DIN |
96 x 96 x P = 95 |
| Available Options: |
| Modelo: |
|
|